Effective membrane filtration for the protection of immunodeficient patients against waterborne bacteria
- Safe water hygiene
- Full working height between faucet and sink
- Operating time up to 2 months
- CE-marked medical device
- Environmentally friendly and efficient reusable product
- Size: L 108 mm x Ø 43 mm
- Retention: 7 log steps Brev. dim.
- Pore size: 0.2 µm
- Operating time*: up to 2 months
- Chlorine retention**: ≤ 10 ppm
- Minimum flow rate: 6 l/min at 5 bar, at 25°C
- Maximum operating pressure***: 5 bar at 60°C
- Maximum operating temperature: 60°C (at 70°C < 10 min. over operating time)
- Adaption: Quick-release coupling
* Depending on the water quality.
** Continuous dosage of ≤ 10 ppm over operating time; short-term (1h) high dosage (400,000 ppm) for chemical disinfection.
*** The Germlyser® A is not suitable for connection to low-pressure systems (e. g. boilers).